I have a problem with my Voluson 730. The volumetric transducer in 2D mode works perfectly, but when I select the option 3D/4D, white cones appear. Is the probe the problem? Thanks.

The probe might be the problem. Swap this probe and test the system to rule out any probe issue. If it is the system, the following items should be considered: Software corruption or the CCM/CRS boards, based on the breakthrough of the system. Voluson 730 systems BT02 and before have a CCM board versus a CRS, like on the newer BTs.