We have a Toshiba Zario ultrasound. After an electricity outage the date has been changed to 01/01/2002, which made a lot of confusion. Please advise how to reset it back to today’s date. Thank you. –
I would locate the CMOS battery (usually a button battery) and measure the voltage. If the voltage is below 3.1V then replace it while the system is plugged in, main switch turned on, and system NOT turned on. Go into setup after the system has powered up and change the date and time to the current date and time.