Hello from Bulgaria. I have a Voluson 730 expert BT05. My system technician replaced the hard drive and reinstalled software. After the restart, it requires an activation key for the ultrasound’s hardware. Where can I find it? –
The permanent key can be found in the options page. Go to utilities, system set up, then options. Also, go to utilities, system set up, and then the system info page to find the hardware ID. These pages should be printed before starting the software reload. If you do not have copies printed already, you have to call the OEM.