I have a Voluson 730 pro that reboots in mid scan. Any ideas on what it could be? –
Check the error logs and the status of the power supply LEDs on the back of the GEF box when the problem occurs. It can be a software corruption. Make sure all patient data has been transferred to the hospital DICOM server, then create a recent system backup including system options. Reload system software according to the procedure in the service manual.