During installation of the system software—version 4.0.5 of the voluson 730 pro GE ultrasound—it gives an error message as follows:
Component Transfer Error
————————-
Components Files
File Group XML
File D:Measure Error Access Is Denied

What does it mean? How do I solve this? Please advise. Thanks and best regards.

The system is unable to write over existing program files due to the nature of the Windows NTFS system. I suggest you ensure you have proper backups of your system according to your user manual and perform a “new disk” option installation.

This function performs operations that result in complete data loss of all files on the C: and D: partition. The image on the system DVD is used for restoring C:. The rescue image stored on R: is replaced by the image from the DVD. Data on D: will be deleted (and re-initialized).

end faq

{accordionfaq faqid=accordion16178 graybarfaq arrow headerbackground headerborder contentbackground contentborder round5}