The U.S. FDA is informing patients, healthcare providers, and manufacturers about a set of cybersecurity vulnerabilities, referred to as “SweynTooth,” that—if exploited—may introduce risks for certain medical devices. SweynTooth affects the wireless communication technology known as Bluetooth Low Energy (BLE).
BLE allows two devices to “pair” and exchange information to perform their intended functions while preserving battery life and can be found in medical devices as well as other devices, such as consumer wearables and Internet of Things (IoT) devices. These cybersecurity vulnerabilities may allow an unauthorized user to wirelessly crash the device, stop it from working, or access device functions normally only available to the authorized user.
To date, the FDA is not aware of any confirmed adverse events related to these vulnerabilities. However, software to exploit these vulnerabilities in certain situations is publicly available. Now, the FDA is providing additional information regarding the source of these vulnerabilities and recommendations for reducing or avoiding risks the vulnerabilities may pose to a variety of medical devices, such as pacemakers, glucose monitors, and ultrasound devices.
“Medical devices are becoming increasingly connected, and connected devices have inherent risks, which make them vulnerable to security breaches. These breaches potentially impact the safety and effectiveness of the device and, if not remedied, may lead to patient harm,” says Suzanne Schwartz, MD, MBA, deputy director of the Office of Strategic Partnerships and Technology Innovation in the FDA’s Center for Devices and Radiological Health.
“The FDA recommends that medical device manufacturers stay alert for cybersecurity vulnerabilities and proactively address them by participating in coordinated disclosure of vulnerabilities as well as providing mitigation strategies,” Schwartz adds. “An essential part of the FDA’s strategy is working with manufacturers, health care delivery organizations, security researchers, other government agencies and patients to address cybersecurity concerns that affect medical devices in order to keep patients safe.”
The FDA is currently aware of several microchip manufacturers that are affected by these vulnerabilities: Texas Instruments, NXP, Cypress, Dialog Semiconductors, Microchip, STMicroelectronics, and Telink Semiconductor. Their microchips may be in a variety of medical devices, such as those that are implanted in or worn by a patient (such as pacemakers, stimulators, blood glucose monitors, and insulin pumps) or larger devices that are in healthcare facilities (such as electrocardiograms, monitors, and diagnostic devices like ultrasound devices).
Medical device manufacturers are already assessing which devices may be affected by SweynTooth and are identifying risk and remediation actions. In addition, several microchip manufacturers have already released patches. For more information about SweynTooth cybersecurity vulnerabilities, including a list of affected devices, see ICS-ALERT-20-063-01 SweynTooth Vulnerabilities, Department of Homeland Security Cybersecurity Infrastructure Security Advisory.