I have a Siemens X300 with boot up problems. First, I found the DC/DC power supply with problems, causing the leads on the USPA-AC power supply to light up. Having repaired the DC/DC power supply, on power up the system hangs on the message “Checking system configurations.” All the voltage tests are OK. What could be the problem? Best regards.

The configuration issue you are experiencing is most likely caused by a hardware compatibility issue. “Checking system configurations” typically indicates that the system is detecting a piece of hardware that was replaced or installed. The DC-DC power supply does not trigger this flag.

The other potential cause is a software issue that a software reload can remedy.