A side-by-side look at defibrillator analyzers, comparing key performance specs, features, and pricing.


Defibrillator analyzers play a critical role in ensuring lifesaving equipment performs as intended when it matters most. For healthcare technology management professionals responsible for inspection, preventive maintenance, and compliance documentation, choosing the right analyzer means balancing accuracy, waveform capability, internal load options, and ease of use—while also considering connectivity, data capture, and CMMS integration. This comparison guide provides a side-by-side look at key specifications, features, and pricing considerations.

BC Group International – DA-2006P


Datrend Systems Inc – Phase 3 Defib Analyzer


Fluke Biomedical – Impulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer


Netech Biomedical – Delta 3300


Rigel Medical – UniPulse 400


Company Name:BC Group InternationalDatrend Systems IncFluke BiomedicalNetech BiomedicalRigel Medical
Defibrillator Analyzer:DA-2006PPhase 3 Defib AnalyzerImpulse 7000DP Defibrillator/Transcutaneous Pacemaker AnalyzerDelta 3300UniPulse 400
Website:bcgroupstore.com/Biomedical-BC_Biomedical_DA-2006P.aspxwww.datrend.com/product/phase-3/flukebiomedical.com/products/defibrillator-analyzers/impulse-7000dp-defibrillator-analyzernetechbiomedical.comrigelmedical.com/product/412a910-rigel-uni-pulse-400/
Description:The DA-2006 is a microprocessor-based analyzer that is used in the testing of defibrillators. It measures the energy output and provides an on-screen view of the defibrillator/pacer pulse. It is used on manual, semi-automatic, and automatic defibrillators with monophasic, biphasic, and pulsed biphasic outputs. The DA-2006 has field upgradability and an optional variable load of 25 – 200 Joules (DA-2006-VL)Phase 3 is a precision defibrillator and pacer analyzer built for modern multiphasic waveforms, combining high accuracy energy measurement with intuitive operation and robust reporting. Designed for reliability, portability, and comprehensive testing, it delivers fast, repeatable verification beyond standard analyzers.The Impulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer is renowned for its superior compliance, technology, efficiency, and user-friendliness. Its rugged exterior houses first-class technology, setting this defibrillator tester apart. The Impulse 7010 Selectable Load Box is a unique, robust, and easy-to-use tool which provides the capability to test defibrillators under a variety of patient impedance conditions to ensure patient safety and is the perfect companion to the Impulse 7000DP.The Delta 3300 is a compact defibrillator and transcutaneous pacemaker analyzer intended for routine performance verification and preventive maintenance. It supports energy measurement, charge time testing, synchronous cardioversion timing, AED analysis, and pacer output evaluation. The unit includes waveform capture with playback capability and integrated ECG and arrhythmia simulation for functional testing. Test records can be stored internally and exported to a PC via RS232/USB for documentation. Its battery-powered design and menu-driven interface support use in both shop and clinical equipment service environments.UniPulse 400 stands out as one of the most portable defib analyzers, combining full AED & ALS testing, high-accuracy ECG simulation, synchronized waveform capture, rapid energy/charge analysis, and built-in data & asset management for efficient PM workflows.
Year Launched:20052009200820192019
Supported Defibrillator Test Functions:Energy output, charge time, ECG simulation, synchronous cardioversion, transcutaneous pacing, AED testing, biphasic waveform analysis, pulsed biphasic, variable loadEnergy output, charge time, ECG simulation, synchronous cardioversion, transcutaneous pacing, AED testing, biphasic waveform analysisEnergy output, charge time, ECG simulation, synchronous cardioversion, transcutaneous pacing, AED testing, biphasic waveform analysisEnergy output, charge time, ECG simulation, synchronous cardioversion, transcutaneous pacing, AED testing, biphasic waveform analysisEnergy output, charge time, ECG simulation, synchronous cardioversion, transcutaneous pacing, AED testing, biphasic waveform analysis
Internal Load/Resistance Options: 50 ohm25 ohm, 50 ohm, 100 ohm, 150 ohm25 ohm, 50 ohm, 100 ohm, 150 ohm, 200 ohm, external load required50 ohm50 ohm
Measurement Accuracy:± 2 % or ± 2 J±1% ±2 LSD0.1 J to 600 J: ± (1 % of reading + 0.1 J)± 1 % or ±  1 J± 1 % or ±  1 J
Supported Standards:IEC 60601-2-4, ANSI/AAMI DF80, IEC 62353IEC 60601-2-4, ANSI/AAMI DF80, IEC 62353IEC 60601-2-4, ANSI/AAMI DF80, IEC 62353n/aIEC 60601-2-4, ANSI/AAMI DF80
Waveform Capability:Live waveform display, waveform capture/storageLive waveform display, waveform capture/storageLive waveform displayWaveform capture/storageLive waveform display, waveform capture/storage, numeric results
Automated Test Sequences/Presets:Yes, user-editable presetsYes, factory presets and user-editable presetsYes, factory presets and user-editable presetsNoYes, factory presets only
Connectivity:Bluetooth, Serial, RS-232USB, BluetoothUSBUSBUSB, Bluetooth
Native CMMS Integration Available:Yes, CMMS integration when using myBC Mobile + myBC Connect.Yes, through vPad PlatformYes, through OneQA Workflow Automation Softwaren/an/a
Battery-Powered:YesYesYesYesYes
Price:$6,585Consult sales for pricingConsult sales for pricing$1,655$4,600-$4,900

Last updated February 2026.

Information for this guide is based on data submitted by defibrillator analyzer marketers. 24×7 strives for accuracy in all data but cannot be held responsible for claims made by marketers. All defibrillator analyzer marketers get one listing at no charge. Advertisers are offered multiple listings. All defibrillator analyzers may not be included. Email [email protected] to be considered for the next update.